Specifications
6-1
December 2004
General Specifications
Accuracy:
0.1 ppm from 0 to 5 ppm
± 2% of reading for all other ranges
Repeatability:
± 0.2% of reading
Reproducibility:
± 0.2% of reading in 24 hours
± 1% of reading in 30 days
Response Time:
less than 1 sec for 90% of step change
Stability:
drift less than 0.1% of sensor output per month
Operating Temperature:
Typical:
700°C
Maximum:
950°C
Ambient Temperature Influence:
no effect
Atmospheric Pressure Influence:
no effect, if sampling gas is vented to atmosphere
otherwise, 0.13% of reading per mm of Hg
Sample Flow Rate Influence:
no effect, if sample flow rate = 1,000 ± 150 cc/min
(2.1 ± 0.3 SCFH)
Line Voltage Influence:
no effect on sensor output
Background Gas Variation Effect:
no effect, if sample gas is clean, dry and free of combustibles
Enclosures:
NEMA 4X/IP66, stainless steel
www.DataSheet4U.com
www.DataSheet4U.com
Содержание CGA 351
Страница 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Страница 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Страница 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Страница 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Страница 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Страница 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Страница 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Страница 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Страница 116: ...CE Mark Compliance Overview B 1 EMC Compliance B 1 LVD Compliance B 1 www DataSheet4U com www DataSheet4U com ...
Страница 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...