Set the O2 flow to 3 l/min.
•
Slowly turn the O2 flow control clockwise, until the N2O flow
is decreased to the rates shown in the following table and
make sure the O2 flow is as specified.
N2O flow l/min
O2 flow greater than l/min
8
2
4
1
0.8
0.2
8.
Adjust the flow of all gases through the full range and make sure
that the flow tubes adjust accordingly.
9.
Disconnect all available O2 pipeline supplies or close the O2
cylinder valve.
10. Make sure that:
•
The low O2 supply alarm occurs.
•
The N2O and the O2 flows stop. The O2 flow should stop
last.
•
The air flow continues (if equipped).
•
The gas supply alarms occur on the ventilator if the
ventilator uses O2 as the drive gas.
•
Turn all the flow controls fully clockwise for minimum flow.
•
Reconnect all available O2 pipeline supplies.
Carestation
™
620/650/650c (A1)
9-10
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...