Auto limits
Use the
Auto Limits
menu to quickly set alarm ranges for ‘MV’, ‘TV’,
and ‘EtCO2’ during mechanical ventilation.
Setting auto limits
1.
Select
Auto Limits
.
The menu shows the current measured values and the
proposed low and high alarm limits.
2.
Check the proposed parameters.
•
Select
Confirm
to use the proposed low and high alarm
limits.
•
Select
Cancel
to leave the alarm limits unchanged.
•
Select
Case Default Limits
to set the alarm limits to the
case default limits.
The proposed low and high alarm limits are shown in highlighted
text. The alarm limits that are not highlighted do not change.
Note
3 Operation
2076152-001
3-15
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...