Water buildup
Water results from exhaled gas and the chemical reaction between
CO2 and the absorbent that takes place within the absorber canister.
At lower fresh gas flows more water builds up because less gas is
scavenged and:
•
More CO2 stays in the absorber to react and produce water.
•
More moist, exhaled gas stays in the patient circuit and the
absorber.
How to help prevent water buildup
Suggested ways to prevent water buildup include:
•
Ensure the water condensing in the breathing circuit tubes is
kept lower than the patient circuit ports and is not allowed to
drain back into the flow sensors.
•
Water condensation in the breathing circuit tubing might be
lessened by using a Heat and Moisture Exchange (HME) filter at
the airway connection.
9 User maintenance
2076152-001
9-13
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...