Setting principal setting defaults
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select the item to change.
For
Principal Volume
, select
MV
or
TV
from the drop-down list.
For
Principal O2
, select
Et
or
Fi
from the drop-down list.
For
Principal Agent
, select
Et
or
Fi
from the drop-down list.
3.
Select
Back
to return to the
System Config.
menu.
Setting volume conditions and humidity
compensation
When set to
ATPD
, the volume calculation conditions are based on
ambient temperature and pressure, dry humidity conditions. When
set to
BTPS
, the volume calculation conditions are based on body
temperature, ambient pressure, saturated humidity conditions.
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select the item to change.
•
To set the volume calculation conditions, set
Vent and
Airway Gas Module TV Basis
. Set to
ATPD
/
ATPD
,
ATPD
/
BTPS
, or
BTPS
/
BTPS
.
•
To set the humidity compensation type, set
CO2 Numbers
to
Dry
or
Wet
.
3.
Select
Back
to return to the
System Config.
menu.
Setting inspiratory flow
1.
From Super User mode, select
System Config.
and then
Parameter Settings
.
2.
Select
Insp Flow
to set the graphical display for flow data.
Select the plus symbol to graphically display inspiratory flow as
a positive value.
Select the minus symbol to graphically display the expiratory
flow as a positive value.
3.
Select
Back
to return to the
System Config.
menu.
Trends setup
Use the
Trends Setup
menu to set the default trend graphical page
layout for the top, middle, and bottom trends for each page shown in
the user’s
Trends
menu.
Note
12 Super user mode
2076152-001
12-7
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...