
Phenomenon
Basic EMC
standard or test
method
Professional healthcare facility environment
Note: UT is the AC mains voltage prior to application of the test level
Recommended separation distances
The system is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer and/or
user of the system can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the system as
recommended below, according to the maximum power of the
communications equipment.
.
RF wireless communication equipment
Test frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity test
level
(V/m)
385
380-390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430-470
GMRS 460,
FRS 460
FM +/- 5 kHz
deviation
1 kHz sine
2
0.3
28
710
745
780
704-787
LTE Band 13,
17
Pulse
modulation
217 Hz
0.2
0.3
9
810
870
930
800-960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2
0.3
28
1720
1845
1970
1700-1990
GSM 1800:
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25: UMTS
Pulse
modulation
217 Hz
2
0.3
28
2450
2400-2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450
LTE Band 7
Pulse
modulation
217 Hz
2
0.3
28
5240
5500
5785
5100-5800
WLAN 802.11
a/n
Pulse
modulation
217 Hz
0.2
0.3
9
11 Specifications and theory of operation
2076152-001
11-43
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...