Case defaults
Use the
Case Defaults
menu to set the default case types that show
on the
Start Case
menu and access the
Volume Apnea Setup
menu.
Configuring case defaults
Set the case defaults for the case types that are selectable from the
Start Case
menu. In each case type, the name, patient defaults, gas
defaults, screen layout defaults, alarm defaults, and ventilator
settings for each available ventilation mode can be preset.
1.
From Super User mode, select
Case Defaults
.
2.
Select the case name to enter the adjustment window.
3.
Select an item to change. Make the change.
4.
Continue to select items and make changes.
5.
Select
Confirm
.
6.
Repeat to set the defaults for another default case.
Setting case name
1.
From Super User mode, select
Case Defaults
.
2.
Select the name of the case to enter the adjustment window.
3.
Select
Name
.
•
Select
Clear
to remove the existing name.
•
Select up to 10 characters from the drop-down list.
•
Select
Delete
to delete a character.
•
Select
Save
to save the name and close the drop-down list.
•
Select
Reset
to return the name to the factory default name.
•
When the 10 character maximum is reached, the name is
automatically saved. The drop-down list closes.
4.
Select
Confirm
when done.
Setting volume apnea defaults
Use the
Volume Apnea Setup
menu to enable or disable the user’s
ability to turn the volume apnea alarms off during manual ventilation.
1.
From Super User mode, select
Case Defaults
.
Carestation
™
620/650/650c (A1)
12-10
2076152-001
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...