9100c
3-4
M1207026
3.2 Inspect the system
CAUTION
The upper shelf weight limit is 25 kg (55 lb).
WARNING
Do not leave gas cylinder valves open if the pipeline
supply is in use. Cylinder supplies could be depleted,
leaving an insufficient reserve supply in case of pipeline
failure.
Before testing the system, ensure that:
•
The equipment is not damaged.
•
Components are correctly attached.
•
The breathing circuit is correctly connected, not damaged.
•
Pipeline gas supplies are connected.
•
Cylinder valves are closed.
•
Models with cylinder supplies have a cylinder wrench attached to
the system.
•
Models with cylinder supplies have a reserve supply of O
2
connected to the machine during system checkout.
•
The casters are not loose and the brakes are set and prevent
movement.
•
The power cord is connected to a wall outlet. The mains indicator
comes on when AC Power is connected.
M1207027
004