9 Repair Procedure
M1207026
9-29
9.12.5 EZchange
Canister spring
replacement
1.
Detach the EZchange module from the machine.
2.
Remove the two M3 shoulder screws (
A
) that fasten the canister
latch lever (
C
).
3.
Remove the latch lever, the switch actuator lever (
B
) and the
spring; discard the spring.
4.
Place the new spring on the module (as shown below). Position
the switch actuator lever over the spring. Ensure the spring hooks
are fully engaged into the posts on the manifold and the actuating
lever.
5.
Clean any residual Loctite debris from the M3 shoulder screws
removed in Step 3.
6.
Place the canister latch lever in position. Apply Loctite 242 to the
threads of the two M3 shoulder screw threads and secure the
canister latch level.
7.
Check the switch actuator lever to ensure free movement. If
sticking is observed, loosen the M3 shoulder screw
approximately 1/8 of a turn until free movement of the switch
actuator lever is observed.
8.
Install the module cover.
9.
Install the EZchange module.
10. Verify that the ‘No CO2 absorption’ message appears on the
screen when the absorber canister is released.
11. Perform the Preoperative Checkout Procedure (refer to the 9100c
User’s Reference manual).
A
B
C
M1207027
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX