9100c
3-24
M1207026
3.13 Electrical safety tests
Make sure the system is completely assembled and all accessory
devices are connected to electrical outlets.
1.
Connect an approved test device (e.g. UL, CSA, or AAMI) and
verify that the leakage current is less than:
2.
Make sure that the resistance to ground is less than 0.2
:
between an exposed metal surface and the ground pin on the
power cord.
Voltage
Max. Leakage Current
100/120 VAC
300
P
Amps
220/240 VAC
500
P
Amps
M1207027
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX