2 Theory of Operation
M1207026
2-9
2.3.2 Manual
ventilation
Manual inspiration
The Bag/Vent switch on the breathing system closes the ventilator
path.
Gas flows from the bag into the breathing circuit module, through the
absorber, and through a unidirectional valve (inspiratory check valve)
to the patient.
During inspiration, common gas flows from the machine into the
inspiratory limb, upstream of the inspiratory check valve.
M1207027
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English−default
Reproduced from the electronic master in MATRIX