Chapter A4: Image Acquisition
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
A4-4
Operator Manual
© 2008 General Electric Company. All rights reserved.
Patient Size:
Allows you to choose the size of the patient being x-rayed.
Available options are:
• Medium Pediatric (toddlers to 5 years old)
• Small Adult
• Medium Adult
• Large Adult
NOTE:
Pediatric techniques are set at different system speeds then
adult techniques. For example, the system speed for a
pediatric exam of 70 kV at 32 mAs is 800. The default system
speed for an adult exam of 70 kV at 32 mAs is 400.
Receptor:
Selects the receptor for the protocol.
In order from left to right, the options are:
• Free Cassette
• GE Portable DR Detector
NOTE:
The generator limits exposure time to two (2) seconds for
both receptors.
Reset Technique:
Resets the technique to the default protocol settings.
Grid position
Shows the recommended grid position for the technique.
SID
Shows the recommended SID for the technique.
kV
Adjusts the kV.
The up/down buttons on the right of the field adjust the kV by one
unit.
The buttons on the left of the field adjust the kV by 5 units.
The kVp selection range is 40-150, in1 kVp increments.
Function
Description