GE MEDICAL SYSTEMS
LOGIQ
200 SERVICE MANUAL
2138853
FUNCTIONAL CHECKS
4–9
REV 0
4–3–2
SMPS Adjustment Procedure
1.
Power LOGIQ
200 ON. Wait for about 30 seconds to warm up the console.
2.
Connect a DVM to the appropriate place shown in Table 4–1.
3.
Verify that the voltages are as shown in Table 4–2.
TABLE 4– 1
SMPS MEASUREMENT LOCATION
SMPS
MEASURE AT
RETURN AT
ADJUST AT
+5V,
±
15V for Analog
V5 (+5V), V–15 (–15V), V15
(+15V) on ESP BD ASSY
TP3 (Ground) on ESP BD
ASSY
1 on SMPS ASSY
See ILLUSTRATION 4–6
–5V for Analog/Digital
V–5 (–5V) on ESP BD ASSY
TP3 (Ground) on ESP BD
ASSY
2 on SMPS ASSY
See ILLUSTRATION 4–6
+12 for Monitor
B+ (+12V) on Monitor Input
Connector
GND on Monitor Input
Connector
3 on SMPS ASSY
See ILLUSTRATION 4–6
+5V for Digital
TP 602 (+5V) on MST BD
ASSY
TP 601 (GND) on MST BD
ASSY
4 on SMPS ASSY
See ILLUSTRATION 4–6
TABLE 4–2
SMPS MEASUREMENT TOLERANCES
SMPS
MIN
MAX
+5V,
±
15V for Analog
+4.8 V
–15.75V
+14.25 V
+5.2 V
–14.25 V
+15.75 V
–5V for Analog/Digital
–4.8 V
–5.2 V
+12 for Monitor
+11.8 V
+12.2 V
+5V for Digital
+4.8 V
+5.2 V
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...