INSTALLATION
2–8
REV 0
2–2
INSTALLATION
2–2–1
Introduction
This section contains many of the procedures required to install the LOGIQ
α
200 console.
2–2–2
Average Installation Time
The LOGIQ
α
200 has been designed to be installed and checked out by an experienced service technician in
approximately three hours. LOGIQ
α
200 consoles with optional equipment may take slightly longer.
2–2–3
Installation Warnings
1.
Since the LOGIQ
α
200 weighs approximately 76 kg (168 lbs) without options, preferably two people should
unpack it. Two people are also preferable for installing any additional bulky items.
2.
There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should
carry out servicing and troubleshooting.
Note
For information regarding packing labels, refer to ILLUSTRATION 2–3, LABELS ON PACKAGE.
3.
After being transported, the unit may be very cold or hot. If this is the case, allow the unit to acclimate before
you turn it on. It requires one hour for each 2.5
°
C increment if it’s temperature is below 10
°
C or above 40
°
C.
CAUTION
Equipment damage possibility. Turning the system on without acclimation after arriving at
site may cause the system to be damaged.
TABLE 2– 2
TIME FOR SETTLEMENT
°
C
60
55
50
45
40
35
30
25
20
15
10
5
0
–5 –10 –15 –20 –25 –30 –35 –40
°
F
140 131 122 113 104
95
86
77
68
59
50
41
32
23
14
5
–4 –13 –22 –31 –40
hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10
12
14
16
18
20
2–2–4
Checking the Components
When a new system arrives, check that any components are not damaged and or missing. If shipping damage or
shortages occur, contact the address shown in Chapter 1.
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...