GE MEDICAL SYSTEMS
2138853
LOGIQ
α
200 SERVICE MANUAL
SYSTEM CONFIGURATION
3–10
REV 3
3–6
TEST POINT, LED, DIP SWITCH AND RESET SWITCH
3–6–1
Test Point List
The table below shows The Test Point list for LOGIQ
α
200.
TABLE 3– 6
TEST POINT
LOCATION
NAME
DESCRIPTION
POSITION
DSC ASSY
TP1
M–mode enable signal
Edge of board
TP3
Mapping clock
–
TP4
Horizontal driving signal
–
TP5
Vertical driving signal
–
TP6
Anti–alasing filter output
–
TP7
Scan line No.0
–
TP8
Scan line No.8
–
TP9
–5V for analog/digital
–
TP10
Input echo signal from the ESP
Assy
–
TP V5_1
+5V for digital
Edge of board
MST ASSY
TP601
Gnd
Edge of board
TP602
+5V for digital
Edge of board
FOOT S/W
Foot switch status
–
CLK2_25M
25MHz clock
–
CLK 9M
9MHz clock
–
CLK36M
36MHz clock
–
RTC ASSY
+5V
+5V for digital
Edge of board
+15V
+15V for analog
Edge of board
–15V
–15V for analog
Edge of board
THV
High voltage for transmitting
Edge of board
RTC_WE
RTC write enable signal
–
CDA
D/A output for continuous
dynamic aperture (CDA)
–
CIM
Not used
–
CDF
D/A output for continuous
dynamic focusing (CDF)
–
PGC
D/A output for Pre–gain
control (PGC)
–
TFC
D/A output for time frequency
control (TFC)
–
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...