GE MEDICAL SYSTEMS
LOGIQ
α
200 SERVICE MANUAL
2138853
6–76
RENEWAL PARTS
REV 5
6–2–34
Printer Bracket Assy(FRU No. 411), Back Bracket Assy (FRU No. 412)
Time Required
30 Minutes
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
Refer to ILLUSTRATION 6–40.
1.
Turn OFF the system.
2.
Remove the Left Cover (FRU 404). Refer to 6–2–27.
3.
Remove the Right Cover(FRU 405). Refer to 6–2–28.
4.
Remove the Front Base Cover (FRU 406). Refer to 6–2–29.
5.
Remove the Front Cover (FRU 407). Refer to 6–2–30.
6.
Remove the Top Cover (FRU 409). Refer to 6–2–32.
7.
Remove the Printer Cover (FRU 410). Refer to 6–2–33.
8.
Disconnect three cables connected to Back Bracket Assy (FRU 412) as listed below. Cut the tie wraps off if
necessary.
a. Unscrew and Disconnect two ring terminals (1 and 2) from the Terminal Block 6 Assy. Refer to 6–2–46,
ILLUSTRATION 6–52.
b. Unscrew and disconnect one ring terminals(Protective Earth) from the chassis.
c. Disconnect two connectors (J5 and J9) from the Mother Assy. Refer to 6–2–22, ILLUSTRATION 6–28.
9.
Remove four hexagonal bolts (1 – 4) and pull out Printer Bracket Assy with Back Bracket Assy.
10.
Unscrew the four screw (5 – 8).
11.
Disassemble Printer Bracket Assy (FRU411) and Back Bracket Assy (FRU412).
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...