6–50
RENEWAL PARTS
REV 5
6–2–20
CONN Assy (FRU No. 306)
Time Required
13 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–26.
1.
Turn OFF the system and plug out.
2.
Remove the Shield Panel (FRU 307). Refer to 6–2–21.
3.
Unscrew four screws (1 – 5).
4.
Disconnect two connector.
5.
Pull out the Probe CONN Assy.
306 CONN Assy
Two Connectors
1
2
3
4
5
CONN ASSY DISASSEMBLY
ILLUSTRATION 6–26
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...