Cleaning
1.
Clean CRT filter.
2.
Clean the Fans by using a vacuum cleaner and bristled brush.
3.
Clean out the interior of the Console by using a vacuum cleaner.
7–2–3
Measurement
1.
Make sure that electrical outlet of hospital should provide the power line standard.
Refer to Chapter 2, Installation.
2.
Ensure all SMPS voltages are within tolerance.
Refer to section 4–3, SMPS ADJUSTMENT in Chapter 4, FUNCTIONAL CHECKS, for adjustment
procedure.
3.
Make sure that leakage currents are within tolerance.
Refer to section 2–3, ELECTRICAL SAFETY TESTS in Chapter 2, INSTALLATION, for measurement
procedures.
7–2–4
Battery replacement
If
Battery Low” message is displayed on the monitor,replace the Battery on MST Assy.
1.
Print out all user preset data for reference.
2.
Remove MST Assy ,slot Number 4 from cardcage.
Refer to 6–2–18, PC BOARD(S) in chapter 6, RENEWAL PARTS for disassembly procedures.
3.
Insert new Battery into empty Battery socket (BAT2) on the edge of MST Assy. Make sure that polarity is
correct (upside po)).
4.
Remove old Battery from socket (BAT1).
5.
Check the user preset data.
CAUTION
If you do not replace Battery even if “Battery Low” message is displayed on the monitor,
SRAM data should be lost and reset to the factory default.
7–2–5
Note
Visual inspection of probes (“head”, cable and connector)
Содержание LOGIQ ?200
Страница 1: ... α ...
Страница 3: ...R D Support Group Of SGMS FAX Number 82 342 420 420 REV 0 2184997 ...
Страница 4: ......
Страница 10: ......
Страница 78: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 2 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 84: ...GE MEDICAL SYSTEMS LOGIQ 200 SERVICE MANUAL 2138853 5 8 DIAGRAMS REV 0 This page is left blank intentionally ...
Страница 101: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 17 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 106: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 22 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 117: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 33 RENEWAL PARTS REV 4 This page is left blank intentionally ...
Страница 135: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 51 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 151: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 67 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 155: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 71 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 159: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 75 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 177: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 93 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 183: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 99 RENEWAL PARTS REV 5 This page is left blank intentionally ...
Страница 194: ...GE MEDICAL SYSTEMS LOGIQ α200 SERVICE MANUAL 2138853 6 110 RENEWAL PARTS REV 5 This page is left blank intentionally ...