Maintenance
• Unit malfunction
• Voided warranty
Use the following procedure to clean and disinfect the cables and leadwires.
1.
Remove cables and leadwires from the device or system before cleaning.
2.
Use care in cleaning leadwires to prevent pulling the long wires from the
connector ends. Metal connections can be pulled away from the connectors.
3.
To clean, wipe with a lightly moistened cloth.
Use a mild soap and water solution to moisten the cloth.
Do NOT use any of the following cleaning products, or products that contain the
same active ingredients and solutions, which are known to cause the problems
previously listed:
• Sani-Cloth® Wipes
• Ascepti® Wipes
• HB Quat®
• Clorox® Wipes (they do not contain bleach)
• Over-the-counter detergents (such as Fantastic®, Tilex®, and so on)
4.
To disinfect, wipe with a soft, lint-free cloth moistened with an appropriate
disinfectant.
Use the following solutions, as recommended in the APIC Guidelines for Selection
and Use of Disinfectants (1996):
• Sodium hypochlorite (5.2% household bleach) minimum 1:500 dilution
(minimum 100 ppm free chlorine) and maximum 1:10 dilution.
• Any sodium hypochlorite wipe product that meets the previous guidelines
can be used.
5.
Observe the following guidelines when cleaning and disinfecting cables and
leadwires.
• Do NOT immerse either end of a cable or leadwire connector.
Immersing or “soaking” the connector ends may corrode metal contact ends
and affect signal quality.
• Do NOT let fluid “pool” around connection pins.
If this happens, blot dry with a soft, lint-free cloth.
6.
Dry with a clean cloth or paper towel.
Cautions
• Follow the cleaning instructions exactly.
• Wring excess disinfectant from wipe before using.
• Never immerse the device, cables, or leadwires in any liquid, as this may corrode
metal contacts and affect signal quality.
• Do not allow fluid to pool around connection pins. If this happens, blot dry with a
soft, lint-free cloth.
• Never use conductive solutions or solutions that contain chlorides, wax, or wax
compounds to clean the device, cables, or leadwires.
158
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......