Introduction
If polarizing electrodes are used, we recommend disconnecting the leadwires from
the patient before delivering the shock.
Electrode defibrillation recovery is the ability of the electrode to allow the ECG
trace to return after defibrillation. We recommend using non-polarizing disposable
electrodes with defibrillation recovery ratings as specified in AAMI EC12 4.2.2.4. (Refer
to the
MAC3500/MAC5500 HD Supplies and Accessories Reference Manual
for a list
of approved electrodes.)
Accuracy Of the Input Signal Reproduction
• Overall System Error meets AAMI EC11 3.2.7.1.
Overall System Error is between or /–5%.
• Frequency Response meets AAMI EC11 3.2.7.2 methods A and D.
Frequency response is between or /- 10% between 0.67 and 40 Hz and
b0 and -10% for 20 ms, 1.5 mV triangular input.
Modulating Effects in Digital Systems
This device uses digital sampling techniques that may produce some variation in
amplitudes of Q, R, and/or S waves from one heart beat to the next, which may
be particularly noticeable in pediatric recordings. If this phenomenon is observed,
the clinician should be aware that the origin of amplitude variations is not entirely
physiologic. For measuring voltages of Q, R, and S waves, it is advisable to use the QRS
complexes with the largest deflection of the particular waves.
Installation and Connection
If the installation of this equipment, in the USA, will use 240 V rather than 120 V, the
source must be a center-tapped, 240 V, single-phase circuit.
Contact GE Healthcare for information before connecting any devices to this
equipment not recommended in this manual.
Supplies and Accessories
For ordering information and a list of supplies and accessories that have been
approved by GE Healthcare for use with this device, see the
MAC 3500/MAC 5500 HD
Resting ECG Analysis Supplies and Accessories Reference Manual
.
Equipment Symbols
The following symbols may appear on the product or its packaging.
Symbol
Description
Type BF equipment. The acquisition module is protected from
defibrillation shocks.
Alternating Current
10
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......