System Setup
Miscellaneous Setup (cont'd.)
Field
Description
Device password
Prevents unauthorized persons from accessing the system. The
password must be entered to use the system.
Enter a 6–character password. By default, the password is blank.
Keep track of all assigned passwords.
SD Card Storage
Only
Determines where the ECGs are automatically stored.
• To store to internal storage, select
No
.
• To store to an SD card, select
Yes
.
NOTE:
Any ECGs that are in internal memory storage will not be
accessible when switching to SD card storage. Be sure that
they have been printed and/or stored before switching to SD
card storage only.
When
Yes
is selected for this option, an SD card must be inserted
in the SD card slot before performing many of the procedures
described in this manual.
Data access speeds may vary, depending on the SD card
capacity and manufacturer. This may affect the time required to
read or write ECG records and other information to the SD card.
For more information on SD cards, refer to
Chapter 13
,
.
Patient Questions
To configure the prompts that are displayed when you select
Patient Data
in the
Resting ECG application, log on to the
System Setup
screen, select
Basic System >
Patient Questions
, and complete the fields described in the following table.
Patient Questions
Field
Description
ID Required
Determines whether you must enter the Patient ID before an ECG
can be recorded.
• Select
Yes
to require the ID.
• Select
No
to make the ID optional.
ID length
Determines the maximum length of the Patient ID. It must match
the same length of the Patient ID in the MUSE CV system to which
the MAC system communicates.
Enter any number from
3
—
16
.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
113
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......