Exercise Stress Testing
• Press the emergency stop button (usually mounted on the treadmill) to stop
the treadmill IMMEDIATELY.
• Press the
Speed W +
,
Speed W –
, and
Grade
keys to manually control the
test. However, once you press these keys, the pre-programmed protocol
becomes inactive: you must manually control the speed and grade during
the remainder of the exercise and recovery phases.
The following menu items are available during the pretest phase of the exercise test:
Menu Item
Description
Patient Data
Enter a patient’s name, ID number, and so forth. Enter the
patient’s age to allow your system to calculate the maximum
and target heart rates.
New Protocol
Select a different exercise test protocol. This function is only
available if the test will be printed only (not stored to memory).
Measurements
Reestablish the median complex, set the J point, and select the
three leads used to calculate heart rate.
Leads
Select the leads used for 3 or 6 Rhythm leads, All Leads, Lead
Check, or Lead Placement.
Median
Select a lead to act as the median lead. This can be a fixed lead
or scanned for lead with most ST depression.
Writer
Change the writer’s arrhythmia documentation (doc.), cubic
spline (baseline control), paper speed, gain, filter, and writer
on/off settings.
Exercise Phase
The selected protocol controls the treadmill or ergometer. When you enter the
exercise phase:
• the belt speed and grade or the ergometer workload changes according to the
selected protocol,
• the exercise clock (top) starts, and
• the system starts to save the test data.
WARNING:
FALLING HAZARD — Severe injury can result from a fall.
Patients should wait until the treadmill belt is moving before stepping onto the
belt. Step onto the belt with one foot at a time. Avoid rapid changes in belt speed.
1.
Press the
Exercise
button to begin the exercise phase.
During the test you can manually perform operations from the function
keyboard.
2.
Press the
Start tmill
button if the treadmill or ergometer has not been started
yet.
If you are using an ergometer, the ergometer workload is automatically
controlled.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
71
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......