Recording an ECG
This opens a list of available gains. Options include
2.5 mm/mV
,
5 mm/mV
,
10
mm/mV
,
20 mm/mV
, and
10/5
. The selected value appears on the function
key on the screen.
• To adjust the filter of the printed ECG, press
150 Hz
.
This opens a list of available filters. Options include
20 Hz
,
40 Hz
,
100 Hz
, and
150 Hz
. This affects only the printout; the screen filter is always set to 40 Hz.
The selected value is displayed on the function key.
• To change the pacemaker enhancement, press
More > Pace Gain
.
This opens the
Pacemaker Enhancement
window. Highlight
Pacemaker
Enhancement
, press
Return
, and select
Yes
.
• To turn on the Acute Coronary Syndrome algorithm, press
More > ACS Off
.
This opens the
ACS On/Off
window. Select
ACS On
and press
Return
.
NOTE:
ACS is an optional feature. This function key is available only if the option
has been purchased and activated. Refer to
“Option Activation” on page
for instructions on activating this option.
NOTE:
ACS is activated on a per-patient basis. Settings are not retained from
patient to patient; it must be activated for each patient.
4.
Once all the settings have been adjusted accordingly, press
ECG
.
Any of the following may occur, depending on your system settings:
• The
ACI-TIPI Required Information
window opens.
This occurs only if the optional ACI-TIPI feature is enabled. If this window
opens, complete the following information:
•
Age (18–40, 41–50, >50)
•
Gender (Male/Female)
•
Chest or Left Arm Pain
• Select
Chief Complaint
if the complaint of chest pain or left arm pain is
the primary reason the patient is at the hospital.
• Select
Secondary Complaint
if chest pain or left arm pain is not the
primary reason the patient came to the hospital.
• Select
Not Present
if the patient is not experiencing any chest pain or
left arm pain or equivalent discomfort.
See
“ECG Acquisition” on page 124
) for information on enabling the optional
ACI-TIPI feature.
NOTE:
If any of the information was provided on the
Patient Information
screen,
you will not be asked for it here.
• The message
Acquiring data x sec
is displayed in the upper left corner of the
screen.
This counts down as it acquires 10 seconds of ECG data.
• A
Critical Test Result
window opens.
This occurs only if (a) the optional Critical Values feature has been enabled
and configured and (b) one or more of the critical values has been met. Press
66
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......