GENERAL INFORMATION
INTRODUCTION
The GDS-68XP provides a safe and reliable way to measure the level of hydrogen sulfide, mercaptan, THT
and other gases in a process stream that contains low levels of oxygen. The GDS-68XP uses a programmed
sequencer to alternatively apply sample and purge air to a highly sensitive and reliable electrochemical
sensor. Applying purge air periodically refreshes the sensor’s chemistry, significantly extends the life of
the sensor and reduces the total amount of sample gas
released into the atmosphere.
The GDS-68XP can be configured to draw samples from
non-pressurized (ambient) sources or to accept positive
pressure samples, from +5 psig to +1500 psig. Multiple
filter options are available and the unit features a low-flow
warning switch and long life brushless DC pump for sample
and purge air.
The GDS-68XP features the new dual-channel GASMAX CX
networked gas monitor. In addition to a new high visibility
color display, the GASMAX CX includes an Ethernet port with built-in MODBUS/TCP interface and web
server for complete integration into user networks. GDS Corp software can be used to remotely monitor
the status and performance of GDS-68XP systems if a network connection is available.
EXPLOSION PROOF INSTALLAT ION
The GDS-68XP is designed for use in Class 1 Division 1 hazardous areas. Installation in these areas should
follow best industry standard practices and all appropriate electrical codes. Generally, these codes require
rigid metal conduit, poured seals and other installation elements necessary to ensure safety. For
maximum protection against RF interference or electrical surge, the GDS-68XP enclosure and
interconnecting conduit must be properly grounded.
INTRINSICALLY SAFE INSTALLATION
The GDS-68XP is not designed or certified for use as an Intrinsically Safe device.