GDS-68XP Operation & Maintenance Manual, Revision 3.4
Page 16
POWER & SIGNAL CONNECTIONS
To access the power and signal wiring, remove the cover from the
explosion proof enclosure. Connect a source of +24VDC, ± 10% power
to wiring terminals J1-1 (+24V) and J1-3 (GND). The 4-20mA (source)
output is available at J1-2 (See Fig. 5-2).
ON-DEMAND SEQUENCE INPUT
If the “on-demand’ sequencer mode is selected the GDS-68XP will enter a hold mode until an external
contact closure between J6-1 and J6-2 occurs. This causes the GDS-68XP to perform a single sample
sequence. Upon completion, the DONE output will be pulled low to signal the remote controller that the
sequence is complete.
Figure 5-2: GDS-68XP Power and Signal Wiring
ETHERNET CONNECTION
The GASMAX CX gas monitor includes a standard RJ-45 Ethernet connection on the main I/O board. Both
fixed IP and DHCP-based dynamic IP addressing is supported. The GASMAX CX supports a full range of
MODBUS / TCP accessible register (see Chapter 10) as well as a built-in web server with data display and
gas detector configuration page.
J1 – 1: +24VDC Power Input
J1 – 2: 4-20mA Analog Out
J1 – 3: Power / Signal Common
Recommended DC Wire Gauge
< 100 ft
#18 GA
100 to 500 ft
#16 GA
500 to 1000 ft #14 GA
J6 – 1: START: Ground to Common to initiate sequence
J6 – 2: Common
J6 – 3: DONE: Open Collector Output, pulled low when sequence complete
Access to optional Alarm Relays,
MODBUS serial ports and Ethernet is
via ¾” NPT fitting on right side of
GASMAX