GDS-68XP Operation & Maintenance Manual, Revision 3.4
Page 49
Ch 2 Balance
43295
3
N/A
Binary
Ch 2 Heater Enable
43296
3
N/A
False = NO, True = YES
Ch 2 Heater Setpoint
43297
3
N/A
Modbus 32 bit IEEE 754 Floating Pt
Ch 2 Temp Comp -40
43299
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp -30
43303
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp -20
43307
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp -10
43311
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp 0
43315
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +10
43319
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +20
43323
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +30
43327
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +40
43331
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +50
43335
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Temp Comp +60
43339
3
N/A
32 bit FP Gain, 32 bit FP Offset
Ch 2 Sensor Type
43343
3
N/A
TBD
Ch 2 Send Sensor Life
43344
3
N/A
False = NO, True = YES
Ch 2 Cal mA Setting
43345
3
N/A
Modbus 32 bit IEEE 754 Floating Pt
Ch 2 Local Cal
43347
3
N/A
False = NO, True = YES
Ch 2 AI Range
43348
3
N/A
TBD
Registers 45001-45022 added in V1.03
Ch 1 Value
45001
3
N/A
800 = “0”, 4000 = Full Scale
Ch 2 Value
45002
3
N/A
800 = “0”, 4000 = Full Scale
Ch 1 Value
45003/04
3
N/A
MODBUS 32 bit floating point
Ch 2 Value
45005/06
3
N/A
MODBUS 32 bit floating point
Ch 1 Alarm 1 Status
45007
3
N/A
“1” = Fault
Ch 1 Alarm 2 Status
45008
3
N/A
“1” = Fault
Ch 1 Alarm 3 Status
45009
3
N/A
“1” = Fault
Ch 1 Fault Status
45010
3
N/A
“1” = Fault
Ch 2 Alarm 1 Status
45011
3
N/A
“1” = Fault
Ch 2 Alarm 2 Status
45012
3
N/A
“1” = Fault
Ch 2 Alarm 3 Status
45013
3
N/A
“1” = Fault
Ch 2 Fault Status
45014
3
N/A
“1” = Fault
Ch 1 Sensor Life
45015
3
N/A
0-100 binary
Ch 2 Sensor Life
45016
3
N/A
0-100 binary
Ch 1 Sensor Temp
45017
3
N/A
Binary 0 - 4095
Ch2 Sensor Temp
45018
3
N/A
Binary 0 - 4095
Ch 1 Sensor Temp
45019/20
3
N/A
MODBUS 32 bit floating point
Ch 2 Sensor Temp
45021/22
3
N/A
MODBUS 32 bit floating point