GDS-68XP Operation & Maintenance Manual, Revision 3.4
Page 30
CALIBRATION PROCEDURE (INLET PORT SAMPLE)
Calibration using the Inlet Port will more closely approximate the
actual sample gas flow path through the filter, regulator and
sequencer valve. Connect the calibration gas cylinder and regulator
to the input port, either directly or via a customer-supplied three-
way valve such that the input can be connected to sample or
calibration gas.
CALIBRATION PROCEDURE – INLET PORT
OBTAIN THE FOLLOWING ITEMS:
1) CYLINDER OF CALIBRATION GAS WITH FIXED FLOW REGULATOR AND LENGTH OF FLEXIBLE TUBING
2) CYLINDER OF ‘ZERO AIR’ IF NECESSARY
IF NOT PREVIOUSLY DONE:
1) VERIFY THAT THE CHANNEL 1
CAL SPAN VALUE
MATCHES THE CALIBRATION GAS TO BE USED
2) VERIFY THAT THE CALIBRATION GAS IS NOT EXPIRED
ENTER CAL MODE:
WAIT UNTIL THE SEQUENCER IS IN PURGE/HOLD MODE
AS
INDICATED BY “L2” (GREEN) FLASH
PRESS THE “CAL” KEY FOLLOWED BY THE “EDIT” KEY
ATTEMPTING TO CALIBRATE THE GASMAX
DURING A SAMPLE SEQUENCE WILL RESULT IN
A CALIBRATION ERROR
THE “APPLY ZERO GAS” SCREEN WILL APPEAR; L2 WILL CHANGE TO SINGLE FLASH
PURGE AIR PUMP WILL TURN ON AND AMBIENT (“ZERO”) AIR WILL FLOW INTO THE SENSOR FLOW CELL
SET THE ZERO VALUE:
ALLOW THE READING TO STABILIZE AND
PRESS THE “EDIT” KEY TO
ACCEPT
THE CURRENT READING AS ‘ZERO’
IF THERE IS A POSSIBILITY THAT TARGET GAS
MAY BE PRESENT IN AMBIENT AIR, USE A
CYLINDER OF ZERO AIR IN PLACE OF PURGE AIR
IF ZERO READING IS WITHIN LIMITS, THE
ZERO CAL SUCCESS
SCREEN WILL MOMENTARILY APPEAR
THE “APPLY SPAN GAS” SCREEN WILL APPEAR
SET THE SPAN VALUE:
CONNECT THE SPAN GAS AS SHOWN ABOVE AND TURN ON THE
REGULATOR.
PRESS “BYPASS” ON THE SEQUENCER
AND VERIFY
FLOW ON THE FLOW METER. ALLOW THE READING TO STABILIZE
AND
PRESS THE “EDIT” KEY TO ACCEPT
THE CURRENT READING AS
‘SPAN’
PRESSING “BYPASS” ALLOWS CAL GAS TO
FLOW THROUGH THE SEQUENCER.
ALLOW THE GAS TO FLOW FOR ABOUT AS
LONG AS IT DOES DURING A NORMAL
SEQUENCE, USUALLY BETWEEN THREE AND
FOUR MINUTES.
IF SPAN READING IS WITHIN LIMITS, THE
SPAN CAL SUCCESS
SCREEN WILL MOMENTARILY APPEAR
THE GASMAX DUAL SCREEN WILL REAPPEAR SHOWING CHANNEL 1 IN PURPLE (CAL DELAY)
COMPLETE CALIBRATION:
PRESS “RUN” ON THE SEQUENCER AND DISCONNECT THE SPAN
GAS CYLINDER
ONCE CAL PURGE DELAY IS COMPLETE, THE
SEQUENCER WILL RETURN TO THE
PURGE/HOLD MODE
Figure 7-2: Calibration Procedure (Inlet Port)