GDS-68XP Operation & Maintenance Manual, Revision 3.4
Page 65
16
SEQUENCER SETTINGS (DIP SWITCH VERSIONS)
Earlier versions of the GDS-68XP sequencer board utilized a DIP SWITCH to set sequence timing values.
For reference, the settings found in earlier units are shown below:
Sequence
Number
DIP SWITCH
Setting (1-2-3-4
1
)
Sample
Time
Purge / Hold
Time
Total Cycle Time
(Approximate)
0
000 1
On-Demand Sequence
30 min
1
2
001 1
1-5 min
5 min
5-7 min
2
2
010 1
1-5 min
10 min
15 min
3
011 1
1-5 min
20 min
30 min
4
3
100 1
1-5 min
50 min
1 hour
5
101 1
1-5 min
110 min
2 hours
6
110 1
1-5 min
230 min
4 hours
7
111 1
1-5 min
350 min
6 hours
NOTE 1
: DIP SWITCH #4 IS USED FOR FACTORY SETUP AND SHOULD ALWAYS BE LEFT
ON
.
NOTE 2
: SEQUENCE 1 & 2 ARE FOR TEST PURPOSES AND SHOULD NOT BE USED WITH ACTUAL SAMPLES
NOTE 3:
RECOMMENDED MINIMUM FOR MOST APPLICATIONS; RECOMMENDED MINIMUM SEQUENCE
FOR MERCAPTANS AND THT IS FOUR HOURS (“1101”)