Inspiration Percent
HAL has a programmable inspiration percent
that allows you to control the inspiratory versus
expiratory ratio. To change this ratio in UNI:
1.
Under the 'Breathing' section in UNI,
locate 'Inspiration Percent'.
NOTE: Inspiration Percent is the percentage
of a breathing cycle where HAL is inhaling.
For example, the default is set at 33%, which
means that HAL will inhale for 33& of the
time and exhale for 67% of the time. When
adjusting inspiration percent also consider
respiratory rate and if the vitals seem
possible to what it is being adjusted to.
O2 Saturation
Use a real oxygen saturation monitor on HAL's left
index finger to get an O2 reading. Prior to using
the oxygen saturation feature, please calibrate the
simulator with the oxygen saturation monitor.
To calibrate the simulator with the
oxygen saturation monitor:
1.
In UNI, click the blue gear in the upper right hand.
O2 Saturation Calibration
2.
Click on 'Inspiration Percent' and adjust the
percentage with the slider bar or by typing in
the numeric value from a range of 11%-90%.