HAL® S1030
|
USER GUIDE
Initial Setup | 11
3.4 INTERNAL BATTERY SETUP
HAL is shipped with a disconnected battery for safety purposes. Before
use, please connect the battery, as shown below.
1.
Locate the battery connectors
on HAL's right-hand side.
2.
Gently lift the right corner of the chest skin
and connect the battery connectors.
WARNING: Do not touch or move the lung bags inside the simulator Doing so will
affect the lung compliance system. If the lung compliance system is affected, it may
have to be returned to Gaumard Repairs or a Field Technician will have to recalibrate
it. Internal components are serviced by Gaumard certified technicians ONLY.
3.
Slide the connected battery wires into the
cavity and secure the skin to HAL's torso.
NOTE: Please charge HAL with the power
supply that came with his accesories. HAL
has a battery run time of approximately 6
hours running on his initial start up settings.
Runtime may vary depending on usage.
Содержание HAL S1030
Страница 2: ...HAL S1030 USER GUIDE ii...
Страница 76: ......