DX4040 INSTRUCTION AND OPERATING MANUAL (E2.04)
Figure 23.
E-mailing options.
6.4.10
Checking Storage Space
Remaining storage space for results and spectra can be verified from Help → Storage
information. An alarm limit can also be set that notifies the user when the storage
threshold has been exceeded (Figure 24).
Figure 24.
Storage information.
6.4.11
Flushing and Switching Off DX4040 via Bluetooth
–
When the measurements are over and after every use of the DX4040, flush
the sample cell properly with instrument nitrogen gas or instrument air.
Clean ambient air can also be used. Sample cell is correctly purged when
the analysis results go back around 0.
–
Check the filter.
–
When batteries are fully charged, brand new and pump is used during
measurements, the DX4040 operates in average more than two and a half hours
depending of the external temperature.
–
Batteries are fully discharged when the DX4040 stops to operate during
measurement. A red LED on the front panel blinks when the state of charge of
59