SLM
[limit]
set soft limit to [limit] %, where [limit] is an integer between 0 and 100
Examples:
slm
query soft limit
90
slm80
decrease soft limit to 80%
Ok
4.6.20
- ALA - Query/reset alarms
Reports the presence of an alarm condition, or resets (clears) the alarm (if possible). A description of the various
alarms can be found in Section 4.7.2. Note that if the condition which created the alarm still exists, clearing the
alarm will cause the unit to immediately issue the alarm again. Also, alarms can be either asynchronous,
meaning that the unit will transmit the alarm string immediately when the alarm condition first arises, or polled,
in that one must issue the
ALA
command regularly to see if an alarm condition has arisen. The choice between
polled and asynchronous alarms is configured with the
ASA
command.
Syntax:
ALA
report current alarm state
ALAR
reset alarm
ALAC
clear alarm (identical to reset alarm)
Examples:
ala
query current alarm state
NONE
no alarm condition
…
…
?A1 – channel power limit
alarm indicating optical feedback is trying to drive a channel past 100%
ala
query alarm state
?A1 – channel power limit
alac
try to clear the alarm
Ok
?A1 – channel power limit
fault condition still exists
pre0
load preset #0
Ok
ala
query alarm state
NONE
RS-7 is properly outputting preset #0
4.6.21
- ASA - Asynchronous Alarm
This command configures whether or not an alarm condition is issued by the unit asynchronously, meaning that
it will transmit the alarm string immediately upon encountering an alarm condition. If is not desirable that the