SpectralLED® RS-7 USER MANUAL
Page 38
4.2.4
- MON – Wavelength Monitor System Query
If the RS-7 is equipped with the optional Wavelength Monitor System, this command requests
centroid/peak/center wavelength (see
PTY
) and (ir)radiance of the currently active wavelength (only a single
wavelength must be active).
Syntax:
MON
query Monitor System
MON C
continuous query (ESC to terminate)
MON F
fast query, executes in under 50 ms
MON CF
continuous fast query (ESC to terminate)
Examples:
scp0,0,15,70
turn off all channels, then set channel #15 to 70% power
Ok
mon
query monitor
475.41,24.51
monitor reports wavelength of 475.41 nm, at an (ir)radiance of 24.5
4.2.5
- LIN - LED Channel Information
Reports centroid/peak/center wavelength (see
PTY
) & FWHM for each channel, in nanometers. Note that this
information is given twice; first is a nominal wavelength & FWHM representing the channel at the current
NPL
Nominal Power Level (default: 70%), and secondly the actual wavelength and FWHM for the channel’s current
operating power level, reported as 0 if the channel is set to off.
Syntax:
LIN
report all channels’ wavelength and FWHM
LIN
[chan]
report wavelength and FWHM for channel [chan]
where [chan] is an integer between 1 and 64.
Examples:
lin
report all channel’s wavelength & FWHM
1,546,11.452,545,11.521
nominal wavelength (
NPL
%), nominal FWHM, actual wavelength, actual FWHM
2,590,23.511,0,0
channel power is set to 0 (off) so actual wavelength & FWHM reported as 0,0
3,615,17.512,616,17.911
…
64,0,0,0,0
unused channels always report 0,0,0,0
final extra <CR><LF> terminates list