SpectralLED® RS-7 USER MANUAL
Page 55
7920.6
outa
7891.9
4.5.5
-
SOB - Standard observer select, 2 degree or 10 degree
Sets or reports the current CIE Tristimulus Standard Observer Included Angle as 2 degrees or 10 degrees. The
default at power-on is always 2 degrees.
Syntax:
SOB
[2|10]
set standard observer to 2 or 10 degrees
SOB
report current Standard Observer
Examples:
SOB
report current Standard Observer setting
10
SOB 2
change Standard Observer to 2 degrees
Ok
4.5.6
-
OXY - Chromaticity of OSP as CIE 1931 x,y
Reports the chromaticity of the current output spectrum
OSP
using the 2 or 10 degree Standard Observer
Included Angle as set by
SOB
.
Syntax:
OXY
report output spectrum
OSP
chromaticity
Returns two ASCII floats separated by a comma indicating the (x,y) coordinate on the 1931 CIE chromaticity
chart.
Examples:
sob2
select 2 degree Standard Observer (not required if 2 degree has been previously set)
Ok
oxy
report chromaticity of output spectrum
OSP
0.3333,0.3333
4.5.7
-
OXYZ - Tristimulus of OSP as CIE X,Y,Z
Reports the tristimulus of the current output spectrum
OSP
using the the 2 or 10 degree Standard Observer
Included Angle as set by
SOB
. Note that for 2 degree observer, the tristimulus Y value will represent the output