Figure 85: Liquid Membrane Separator Cover with Membrane Side Down
9.
Press down on the cover and apply firm pressure to ensure the O-ring is seated properly in
the groove.
10.
Screw the cover back into the body of the separator until it reaches a mechanical stop. Do
not use a wrench or any other tool at this point as this may overtighten and damage the
separator.
11.
Maintenance should be complete, restore the sample flow and pressure to the system.