Figure 56: Location of Bolts Securing Herriott Cell
13.
Remove the bolts using an appropriate size wrench. The Herriot cell should now be free of
any attachments to the component rail.
14.
Lift the Herriott cell off of the component rail and transfer it to a clean workspace to complete
any further maintenance.
15.
To reattach the Herriott cell once maintenance is complete, simply perform all the steps in
this section in the reverse order.
5.3.4 Removing the Herriott Cell from the AccuLase-GPA™ D2
1.
Gain access to the main enclosure by turning the door locks counter-clockwise with a large
flathead screwdriver. See Figure 65.
Figure 57: Location of Main Enclosure Door Locks
2.
Locate the fittings connecting the Herriott cell to the rest of the analyzer. See Figure 66.