40
Table 10: Analysis Parameters
Parameter
Explanation
Unit
Changes the measurement unit displayed for the concentration in the web GUI
and on the analyzer display. NOTICE: Display only. No unit conversion is done.
Default is ppm.
Purge Time
Changes the amount of time, in seconds, that the analyzer will purge the sample
cell after a stream switch before starting a new analysis. Default is 70s.
Sample Interval
Changes the amount of time, in seconds, the concentration measurement is
averaged over before updating the analyzer display and outputs. Default is 20s.
Contract Start Hour
Allows the user to choose what time, from 0 (midnight) to 23 (11pm), analyses
scheduled to be run daily will be carried out.
4.4.4.4 Bypass
The Bypass section contains parameters that govern the behaviour of the AccuLase-GPA™ while
in bypass mode. In bypass mode, hardware outputs associated with alarm conditions are disabled,
and the analog output(s) is/are set to output the fail-safe value (3mA). If the Timeout Enable
checkbox is selected, the analyzer will automatically exit bypass mode after a specified period of
time with no inputs being made to the analyzer either from the keypad or from the web GUI. The
bypass timeout length, in seconds, is user configurable in the Timeout (seconds) field. The default
value for this timeout is 240 seconds (4 minutes).
4.4.4.5 Laser
The Laser section contains several laser-related parameters. These parameters are read-only and
are primarily used for troubleshooting purposes. The parameters displayed in the Laser section
are the same as those displayed on the Laser panel of the AccuLase-GPA™ local display. Refer to
section 3.9 for more information.
4.4.5 Calibration Page
The Calibration page shows the AccuLase-GPA™ calibration parameters. See Figure 37.
Figure 29: Calibration Page
The parameters displayed on the Calibration page are identical to those displayed on the
Calibration panel of the analyzer’s local display UI. Refer to section 3.8 for more information on
these parameters. Unlike the local display Calibration panel, the calibration parameters displayed