Revision 1
17
6/27/2017
2.2.4 Instrument Air
The system requires 22 SCFM at 50-100 psi.
2.2.5 Steam
Two supplies are required:
•
Nozzle: Pressure = 40-80 psi, maximum temperature = 160
o
C
•
Oven Heater: Pressure = 140-175 psi, maximum temperature = 200
o
C
2.2.6 Area Classification
2.3 Unpacking
The 943-TGS analyzer system is packed for shipment in a wooden crate.
To unpack the system:
a) Remove the lid by undoing the marked lag bolts.
b) Once the lid is off, remove the excess packing material, boxes and sample probe from
the shipping crate. The sample probe will be wrapped in packing material. Probes with a
guide length of 5.5 feet or longer are packed in a separate crate.
c) Visually inspect the small packages and the sample probe to ensure that no major
damage has occurred. If damage has occurred, contact the shipping company and
Galvanic Applied Sciences. Place the small packages and the sample probe aside in a
safe, secure storage area as they are not needed at this stage of the system installation.
The control cabinet weighs approximately 62 kg
and the oven cabinet
weighs approximately 60 kg. Use suitable precautions when removing it
from the crate and placing it in the facility.
d) Locate and remove the four (4) bolts that secure the control cabinet framework to the 2' x
4' boards at the bottom of the shipping crate.
e) Once the bolts are removed the control cabinet can be lifted from the crate.
943-TGS-CE
Device Group:
II 2 G.
Ex protection:
II 2 G Ex d e ia mb px IIC T3 Gb.
Ambient Temperature: -20°C to +60°C.
943-TGS-CE
Steam is supplied to the nozzle only. There is no requirement for oven heater
steam on the 943-TGS-CE. There is no steam heater in 943-TGS-CE oven.
Содержание 943-TGS
Страница 23: ...Revision 1 23 6 27 2017 Figure 2 3 Potted Fixture Figure 2 4 Control Cabinet Fiber Optics Entries ...
Страница 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Страница 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Страница 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Страница 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Страница 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Страница 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Страница 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Страница 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Страница 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...