calibration or fault condition exists, the analyzer automatically holds the last valid Air Demand
output value until normal operation is re-established.
The
analog outputs
are updated at a rate of once per second, driven by the calculation cycle
frequency.
3.3.2 Digital Outputs
There are four (4)
digital outputs
available from the analyzer. The
digital outputs
are present
in the form of zero potential form C relay contacts. The
digital outputs
are isolated to 250
V
rms
. The
digital outputs
are provided to indicate analyzer operational status. The operational
status outputs are:
•
Status (Fault / Normal)
•
Service (Warning / Normal)
•
Mode (Calibrate / Run)
•
Control (Manual / Auto).
In addition to the 4 digital outputs associated with the analyzer operation, there is an
independent Purge Fail Alarm attached to the purge system.
a) Status (Fault / Normal) Indicator
The
Status (Fault / Normal)
output provides analyzer system readiness information.
Before ‘Normal’ status indication is given, the analyzer temperature zones and
spectrometer signal must be within the specific set point ranges. This indicator will be in
the ‘Normal’ state
only when ALL
of the following are true:
•
the analyzer is receiving mains power.
•
the cell temperature is between the cell ‘Low Alarm’ temperature and ‘High Alarm’
temperature set points.
•
the probe temperature is between the probe ‘Low Alarm’ and cell ‘High Alarm’
temperature setpoints or the sulphur vapour signal is less than the High Svap alarm.
•
the last spectrometer ‘Measure’ scan was within range.
•
the last spectrometer ‘Reference’ scan was within range.
•
the last spectrometer ‘Dark’ scan was within range.
The purpose of this indicator is to signify an analyzer problem. When this indicator is in
the ‘Fault’ state, the analyzer reverts to a ‘Back purge’ state. Should the fault causing
criterion clear, the analyzer will automatically return to ‘Normal’ status and start sampling.
b) Service (Warning / Normal) Indicator
The
Service (Warning / Normal)
output indicates when service is requested. This
indicator will be in the ‘Normal’ state
unless one or more
of the following is true:
•
the control cabinet temperature is below 15ºC or over 40ºC.
•
the spectrometer integration period is greater than 500 ms.
•
the peak value of the absorbance vector is greater than 2.0.
•
the sulfur vapour concentration is greater than 0.2%.
•
the peak value of the raw scan during last zero is less than 2500 A/D (analog to digital)
counts.
Содержание 943-TGS
Страница 23: ...Revision 1 23 6 27 2017 Figure 2 3 Potted Fixture Figure 2 4 Control Cabinet Fiber Optics Entries ...
Страница 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Страница 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Страница 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Страница 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Страница 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Страница 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Страница 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Страница 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Страница 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...