Reminders to Use the Equipment Safely
Precautions when Using with Wireless LAN Adapter
When using a wireless LAN, make sure to check the operation of any medical electrical equipments nearby.
Stop using the wireless LAN immediately if unexpected waveform noise occurs or there is interference from
surrounding equipment.
Precautions Concerning the Wireless LAN Adapter
Use a wireless LAN adapter specified by Fukuda Denshi.
If an unspecified wireless LAN is used, not only that the communication failure with the DMS may occur, but
the damage of the main unit may also occur. The warranty does not cover data retrieval or repair that is required
when problems such as these are caused by using unspecified items.
Precautions about the Static Electricity
The wireless LAN adapter is extremely delicate and is very vulnerable to damage from electrostatic discharge.
If the human body comes in contact with the metal part of the wireless LAN adapter, static electricity may be
discharged onto the parts or into the circuit of the wireless LAN adapter, which may damage the wireless LAN
adapter. Make sure to follow the precautions below to protect the wireless LAN adapter and this equipment
from damage caused by static electricity.
Remove static electricity from the body before using.
Hold the end of the wireless LAN adapter when handling it. Never touch the metal part of the wireless LAN
adapter.
Storage, Transportation, and Disposal
Storage and Cleaning
CAUTION
Do not open the housing.
Wipe off dirt on the surface of the housing with a soft cloth moistened with ethanol or with
the optional silicon cloth.
Perform daily checks to maintain safety.
Perform regular inspections every 12 months to maintain safety.
Check all cables, equipments, accessories, earth impedance, leakage current, and
accuracy.
If the device has not been used for a while, make sure to check that the device operates
properly and safely before using it.
Be careful not to let liquid such as alcohol get into the equipment.
Transportation
CAUTION
Pack this device in packaging specified by Fukuda Denshi when transporting.
The device does not have handles. Carry it by holding it with both hands.
CAUTION
CAUTION
CAUTION
bl5652
Содержание CardiMax 2
Страница 4: ...2 ...
Страница 10: ...Contents ...
Страница 26: ...Chapter 1 General Description Contents ...
Страница 34: ...Chapter 2 Operational Description Contents ...
Страница 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Страница 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Страница 48: ...Chapter 4 Spare Parts List Contents ...
Страница 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Страница 67: ...6 1 Chapter 6 Software Update Contents Chapter 6 Software Update Software Updating Procedure 6 1 ...
Страница 68: ...Chapter 6 Software Update Contents ...
Страница 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Страница 72: ...Chapter 7 Troubleshooting Contents ...
Страница 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Страница 82: ...Chapter 8 Maintenance Contents ...
Страница 104: ...Chapter 9 Maintenance and Inspection Contents ...
Страница 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Страница 116: ...Chapter 10 Appendix Contents ...
Страница 122: ...10 6 Chapter 10 Appendix Check List ...
Страница 123: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...
Страница 124: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101440 201706 ...