9-5
Chapter 9 Maintenance and Inspection
Electrical Safety Inspection Procedure
Other
Electrical Safety Inspection Procedure
Electrical safety-related testing methods and measurement equipment are stipulated in safety test standards (EN
60601-1). A dedicated safety test measuring instrument is required for performing this measurement . Check that the
measurement results do not exceed the threshold value and record the numerical values.
Safety Evaluation Test Device
Testing and inspection items in this manual are described based on the assumption that a dedicated electrical safety
tester with its accuracy controlled is used. Considering ease of test and ensuring accuracy, it is recommended to use
a dedicated safety test measuring instrument.
Contact our service personnel for more information on safety test measuring instrument.
Check Items
Electrical safety check items include the following types.
Measurement is performed under the following two conditions.
Item
Inspection Item
Procedure
Criteria
Other
1) Power supply
Check whether the device can be turned
on/off both with AC power and battery
power.
The device should be turned on/off both
with AC power and battery power.
2) Power cable
Perform continuity test using a tester.
Perform a continuity test using a tester
and check that the equipment is
conducting.
(1)
Earth Leakage Current
Leakage current that runs to the protective earth conductor.
(2)
Touch Current
Leakage current that runs through the ground from the enclosure of the device
(3)
Patient Leakage Current
(From patient connection to earth)
Leakage current that runs through the ground via patient from the attaching part
There are two types of measuring method - direct current and alternating current.
(4)
Patient Leakage Current
(Current when external voltage is
applied to SIP/SOP)
Leakage current that runs through the ground via patient from the attaching part by
means of the supply voltage carried on the signal input/output
There are two types of measuring method - direct current and alternating current.
(5)
Patient Leakage Current
(When external voltage is applied to
patient connection of Type F Applied
Part)
Leakage current that runs through the ground from the device by means of the supply
voltage carried on the F type attaching part through patient
(6)
Patient Auxiliary Current
Current that runs between attaching parts through patient
There are two types of measuring method - direct current and alternating current.
bl5867
(1)
Normal Condition
Measurement in normal condition
(2)
Single Failure Condition
Condition under which a single fault occurs
Содержание CardiMax 2
Страница 4: ...2 ...
Страница 10: ...Contents ...
Страница 26: ...Chapter 1 General Description Contents ...
Страница 34: ...Chapter 2 Operational Description Contents ...
Страница 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Страница 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Страница 48: ...Chapter 4 Spare Parts List Contents ...
Страница 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Страница 67: ...6 1 Chapter 6 Software Update Contents Chapter 6 Software Update Software Updating Procedure 6 1 ...
Страница 68: ...Chapter 6 Software Update Contents ...
Страница 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Страница 72: ...Chapter 7 Troubleshooting Contents ...
Страница 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Страница 82: ...Chapter 8 Maintenance Contents ...
Страница 104: ...Chapter 9 Maintenance and Inspection Contents ...
Страница 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Страница 116: ...Chapter 10 Appendix Contents ...
Страница 122: ...10 6 Chapter 10 Appendix Check List ...
Страница 123: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...
Страница 124: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101440 201706 ...