Preface
Reminders to Use the Equipment Safely
Do not ground with a gas pipe. Explosion or fire may result.
MRI
Before performing MRI scans, disconnect the electrodes connected to the patient.
Local heating due to induced electromotive force may cause burns to the patient. For details, refer to the
operation manual for the MRI testing device.
Defibrillator
When using this equipment with a defibrillator, use only the specified lead cable. The use of a lead cable other
than the specified defibrillation-proof type may damage the equipment or compromise safety when the
equipment is used along with a defibrillator.
Do not touch this equipment or patient during defibrillation. It may cause an electric shock.
While using a defibrillator, ECG cannot be displayed or recorded.
High-frequency Surgical Devices
When using this equipment with a high frequency external surgical device (electrosurgical instrument), make
sure the contact between the patient and the ground plate is secured. If the connection is incomplete, the patient
may suffer a burn at the electrode site.
When using with electrosurgical instruments, make sure the contact between the patient and the ground plate
can be sufficiently secured. If the connection is incomplete, the current from the electrosurgical instrument may
run into the electrodes of this equipment and the patient may suffer a burn at the electrode site. The electrode
should also be attached as far away from the surgery site as possible to decrease the risk of burns.
The amount of noise interference is considerably different depending on the electrode position and surgery site.
Place the ECG electrodes as far away as possible from the surgery site and the ground plate.
Plug the electrosurgical instrument into a hospital grade outlet as far away as possible from the outlet of this
equipment. This will reduce interference from the electrosurgical instrument.
While using a high frequency surgical device, ECG cannot be displayed or recorded.
Other Equipments
Do not connect any unauthorized equipments.
Connect so that all equipments satisfy the conditions of operating environment.
Only the peripheral devices and cables specified by Fukuda Denshi should be connected to this equipment
using the specified procedure. For details of the connection procedure, contact your local Fukuda Denshi
service representative.
Contact your local Fukuda Denshi service representative when connecting with other equipments. The
connecting equipment must comply with the IEC or ISO standards, or equivalent safety level. Otherwise, it
may cause hazardous situation such as electric shock to the patient and operator.
When using with other equipments, the environment conditions (including transport/storage) of the connecting
equipments should be also observed. Failure to do so may damage the equipment. For the environment
conditions, refer to the operation manual of each equipment.
The operator should not touch the patient and the system composed of this equipment at the same time. It may
cause hazardous situation such as electric shock to the patient and operator.
Optional Accessories
WARNING
Use only the cables specified by Fukuda Denshi.Use of other cables may result in increase
in emission or decrease in immunity.
WARNING
WARNING
WARNING
WARNING
bl5641
Содержание CardiMax 2
Страница 4: ...2 ...
Страница 10: ...Contents ...
Страница 26: ...Chapter 1 General Description Contents ...
Страница 34: ...Chapter 2 Operational Description Contents ...
Страница 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Страница 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Страница 48: ...Chapter 4 Spare Parts List Contents ...
Страница 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Страница 67: ...6 1 Chapter 6 Software Update Contents Chapter 6 Software Update Software Updating Procedure 6 1 ...
Страница 68: ...Chapter 6 Software Update Contents ...
Страница 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Страница 72: ...Chapter 7 Troubleshooting Contents ...
Страница 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Страница 82: ...Chapter 8 Maintenance Contents ...
Страница 104: ...Chapter 9 Maintenance and Inspection Contents ...
Страница 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Страница 116: ...Chapter 10 Appendix Contents ...
Страница 122: ...10 6 Chapter 10 Appendix Check List ...
Страница 123: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...
Страница 124: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101440 201706 ...