FUSION SPLICER 90S Rev5
71
T
he dimensions of the Protection Sleeve after heat shrink
Form
Tension
member
Sleeve
length
Prepare fiber
length
Diameter of
an adaptation
optical fiber
Diameter
of a result
FP-03
SUS
60mm
16mm or less 250~900um 3.1mm
FP-03(40mm)
SUS
40mm
10mm or less 250~900um 3.1mm
FP-04T
Glass Ceramic 40mm
10mm or less 250~900um 4.0mm
FPS01-400-15 SUS
15mm
5mm or less
~400um
1.5mm
FPS01-900-20 SUS
20mm
6mm or less
~900um
2.3mm
The dimensions of the protection sleeve after shrink vary depending on
the diameter of the fiber.
Referring to or Editing Heater Mode
Tube-heating conditions stored in heater mode can be edited or changed
1.
Select the icon in the [Heater Mode] menu, to show the heater
mode.
The settings category is displayed on the left side. The parameters
included in the category are displayed on the right side.
2.
Select and change the category icon and find the target parameter.
3.
Select the parameter displayed on the right side to show the parameter
edit screen.
4.
Select and change the target parameter.
Sleeve length
*
Outer tube
Inner tube
Tension member