FUSION SPLICER 90S Rev5
99
Error Message List (7 of 8)
Error Message Reason
Solution
High
Estimated
Loss
(2/2)
・
Inadequate arc power.
・
Calibrate the arc power with
the [Arc Calibration] function.
・
Using unsuitable splice
mode
・
Select a suitable splice mode
for the fibers to be spliced.
・
[Loss Limit] is set too low.
・
Increase [Loss Limit] to an
adequate limit.
・
Inadequate arc
parameters in special
mode
・
Confirm the arc parameters
are adequate to splice the
fibers.
・
Inadequate estimating
parameters in special
mode
・
Confirm the estimating
parameters are adequate to
estimate the loss. The MFD
mismatch function does not
work for certain types of
specialty fibers. In these
cases, set the [MFD
Mismatch] to “OFF”.
Error Message List (8 of 8)
Error Message
Reason
Solution
There is dust
after executing
Dust Check
function
・
Dirt or dust exists in
optical path.
・
Clean the objective lenses by
referring see section
[Cleaning of Objective Lens].
・
When the above processes
cannot remove the dirt or
dust, consult the nearest
Fujikura authorized
distributor.
X Camera
Y Camera
Trouble
・
The Camera may be
damaged.
・
Consult the nearest Fujikura
authorized distributor.
Heater Oven
Trouble
・
Heater does not heat.
・
Consult the nearest Fujikura
authorized distributor.
Communication
error
・
Broken the internal
device
・
Consult the nearest Fujikura
authorized distributor.
Temperature
Sensor NG
・
Temperature Sensor
may be damaged.
・
Consult the nearest Fujikura
authorized distributor.
Cover Trouble
・
Cannot Open and Close
cover
・
Check sheath clamps are
closed.
・
Confirm obstructions and
fibers are not around the
cover.
・
Confirm sheath clamps or