![FujiFilm PB2020-M Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/fujifilm/pb2020-m/pb2020-m_operation-manual_2343363056.webp)
Electromagnetic immunity compliance information and guidance
Immunity test
EN 60601-1-2
Test level
Compliance
level
Guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
± 6kV: contact
± 8kV: air
Same as
left
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast transient/
burst
EN 61000-4-4
± 2kV: for power supply
lines
± 1kV: for input/output lines
Same as
left
Main power quality should be
that of a typical commercial or
hospital.
Surge
EN 61000-4-5
± 1kV: Line to line
± 2kV: Line to ground
Same as
left
Main power quality should be
that of a typical commercial or
hospital.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
< 11.5V (> 218.5V dip)
For 0.5 cycle
92V (138V dip)
For 5 cycle
161V (69V dip)
For 25 cycle
< 11.5V (> 218.5V dip)
For 5 sec
Same as
left
Main power quality should be
that of a
typical commercial or hospital. If
the user of this product requires
continued
operation during power mains
interruptions, it is recommended
that this product is powered from
an uninterruptible power supply
or battery.
Power frequency (50/60
Hz) magnetic field
EN 61000-4-8
3 A/m
Same as
left
It is recommended to use this
product by maintaining enough
distance from any equipment that
operates with high current.
202B1263042
56
Main Specification
PB2020-M_E2-10_202B1263042.indb 56
2016/03/16 16:43:40