5.5.2
Rinsing of Endoscopic Ultrasonic Probe
WARNING
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, disinfection to be performed at the next step
will lose its effectiveness.
CAUTION
After cleaning, rinse off any remaining detergent solution with water. If
detergent solution remains, chemicals to be used in the subsequent steps
may be diluted or adulterated.
After cleaning, rinse off any remaining detergent solution with clean water.
If detergent solution remains, it may flow into the patient's body.
[Note] Perform all rinsing steps while the endoscopic ultrasonic probe is fully
immersed in clean water.
(1)
With fresh clean water, fill a basin of adequate
size to completely immerse the endoscopic
ultrasonic probe.
(2) Fully immerse the endoscopic ultrasonic probe
in clean water, and then rinse it by gently
shaking it.
(3) Wipe the external surface (insertion portion,
boot, probe connector, end holder and
connector cap) of the endoscopic ultrasonic
probe with clean lint-free gauze in clean water.