(4) Press the [FREEZE] key on the keyboard of the
ultrasonic processor to disable freeze mode.
Make sure that the transducer of the ultrasound
head is rotating and that a multiple-echo image
appears on the monitor.
[Note]
Make also sure that the insertion portion and
other components function normally and do not
produce abnormal sound.
(5) Make sure that the image on the monitor
shakes when shaking the endoscopic ultrasonic
probe's distal end in sterile water.
(6) Press the [FREEZE] key on the keyboard of the
ultrasonic processor to enable freeze mode.
[Note]
Make sure that the transducer of the ultrasound
head stops rotating.
202B1263042
28
Chapter 3 Inspection Before Use