5.5 Manual Cleaning
Perform manual cleaning in a sink, etc. after pre-cleaning.
Manual cleaning must be performed immediately after pre-cleaning is completed.
5.5.1
Cleaning of Endoscopic Ultrasonic Probe
(1) Attach the connector cap to the probe
connector.
(2) Fill a clean basin with detergent solution at the
temperature and concentration recommended
by the detergent solution’s manufacturer.
(3) Fully immerse the endoscopic ultrasonic probe
in detergent solution.
[Note]
Perform all cleaning steps while the endoscopic
ultrasonic probe is fully immersed in detergent
solution.
(4) Brush the probe connector, the connector
cap and the boot with the toothbrush for 20
seconds.
202B1263042
39
Chapter 5 Cleaning