(4) Insert the endoscopic ultrasonic probe into the
forceps inlet of an endoscope with the forceps
valve on.
[Note]
Open the lid of the forceps valve before
insertion.
(5) Insert the endoscopic ultrasonic probe up to a
position where the ultrasound head is visible in
the endoscopic image.
(6) Press the [FREEZE] key to enable freeze
mode. "SCAN ON" is displayed at the lower
right of the monitor, and the transducer starts
rotating. The radial image will be displayed on
the monitor.
(7) In radial scanning, ultrasound beam is radiated
with the endoscopic ultrasonic probe as its
center. In order to expose the target region
perpendicular to ultrasound beam, angulate
the endoscope so that the ultrasound head is
parallel to the target region while monitoring
the endoscopic image on the monitor.
A radial image is displayed on the monitor in
real time.