![FujiFilm PB2020-M Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/fujifilm/pb2020-m/pb2020-m_operation-manual_2343363044.webp)
6.2 Disinfectant Solution
CAUTION
Do not immerse the endoscopic ultrasonic probe in the ozone water.
Doing so may damage the endoscopic ultrasonic probe.
The material compatibility of the endoscopic ultrasonic probe with legally marketed 3.0
to 3.5% glutaraldehyde disinfectant solutions has been verified. Use glutaraldehyde
disinfectant solutions in accordance with the manufacturer’s instructions. For the
names of specific brands of disinfectant solution, contact your local FUJIFILM dealer.
[Note] Refer to the disinfectant solution manufacturer’s instructions regarding
preparation, use, exposure conditions and expiration date.
[Note] Check the disinfectant solution’s efficacy (Minimum Effective Concentration)
before use according to the manufacturer’s recommendation (such as using a
test strip), referring to the disinfectant solution manufacturer's instructions.
202B1263042
44
Chapter 6 Chemical Disinfection
PB2020-M_E2-10_202B1263042.indb 44
2016/03/16 16:43:39